Stimulus : CENTER set near centerfrequency of crystal,
Stimulus : SPAN big enough to see +45 dgr and - 45 dgr phase shift.
Traces : CH1 LOGMAG and CH1 PHASE
Calibration : full (open, short, load, isoln, thru)"
Usage
Step 1 :
Setup NanoVNA as mentioned above.
Step 2 :
Use marker to search for CH1 Phase close as possible to 0°
Register fR (frequency at 0°) in example screenshot below : 15996181 Hz
and S21 LogMag (dB at 0°) in example screenshot below : -2.09 dB
Step 4 :
Use marker to search for CH1 Phase close as possible to +45°
Register fL (frequency at +45°) in example screenshot below : 15995825 Hz
Step 5 :
Use marker to search for CH1 Phase close as possible to -45°
Register fH (frequency at -45°) in example screenshot below : 15995688 Hz